Sonatest D70 Digital Ultrasonic Flaw Detector
The Sonatest Masterscan D-70 Digital Ultrasonic Flaw Detector is a professional NDT instrument designed for internal defect detection, weld inspection, and ultrasonic thickness measurement. Featuring a 450V high-voltage pulser, seven measurement modes, DAC technology
Description
Sonatest D70 Digital Ultrasonic Flaw Detector – Professional Non-Destructive Testing Equipment
The Sonatest Masterscan D-70 Digital Ultrasonic Flaw Detector is a high-performance non-destructive testing (NDT) instrument designed for detecting internal material discontinuities, inspecting weld integrity, measuring material thickness, and evaluating structural conditions without damaging the test object.
Developed by Sonatest, the Masterscan D-70 combines advanced ultrasonic signal processing, flexible inspection configurations, and specialized defect-sizing tools in a portable industrial instrument. It is designed to support professional inspectors working in manufacturing, welding, oil and gas, power generation, aerospace, and industrial maintenance environments.
The D-70 offers a comprehensive range of ultrasonic inspection capabilities, including pulse-echo testing, transmit-receive testing, angle-beam inspection, thickness measurement, and time-of-flight evaluation. Its high-voltage pulser, selectable frequency filters, and advanced signal processing functions provide flexibility when examining different materials and component geometries.
A major advantage of the Masterscan D-70 is its expandable software architecture. Standard inspection functions can be supplemented with optional technologies such as Split DAC, TCG, AVG/DGS, AWS, API, corrosion thickness logging, and Dryscan functionality, depending on the selected configuration.
With encoded B-Scan capability, advanced thickness logging, onboard data storage, and UT-Lity data management software, the Sonatest D-70 provides a practical solution for inspection professionals who require reliable ultrasonic examination and organized inspection documentation.
Key Features of Sonatest Masterscan D-70
- Advanced Digital Ultrasonic Testing: Professional ultrasonic flaw detection and thickness measurement in one instrument.
- High-Voltage Pulser: Adjustable 100–450 V negative spike and square-wave pulser.
- Selectable Frequency Filters: Eight selectable frequency bands for different ultrasonic inspection requirements.
- Seven Measurement Modes: Multiple measurement configurations for flaw detection, depth evaluation, and thickness gauging.
- Dual Inspection Gates: Two independent gates for monitoring and evaluating ultrasonic signals.
- DAC Function: Standard Distance Amplitude Correction for reference-based indication evaluation.
- Advanced Defect Sizing: Optional Split DAC, TCG, AVG/DGS, AWS, and API functions.
- Active Edge Technology: Specialized near-surface resolution enhancement.
- Automatic Calibration: AutoCal calculates material velocity and probe zero using two reference echoes.
- Encoded B-Scan: Supports encoded scanning and cross-sectional inspection data.
- Advanced Thickness Logging: Supports corrosion monitoring and thickness data management.
- Rugged Portable Design: Impact-resistant enclosure for industrial inspection environments.
- Extended Battery Operation: Up to 10.5 hours according to the current manufacturer’s product page.
- PC Data Management: USB connectivity and UT-Lity software for inspection data organization.
Advanced Ultrasonic Inspection Technology
Digital Ultrasonic Flaw Detection
The Sonatest D-70 operates by transmitting high-frequency ultrasonic pulses into a test material through an ultrasonic transducer. The instrument receives reflected signals generated by material boundaries, internal discontinuities, and other acoustic interfaces.
By analyzing signal amplitude, travel time, and waveform characteristics, inspectors can identify potential internal defects and determine their approximate positions.
The instrument supports different inspection techniques to accommodate a wide variety of industrial materials and inspection requirements.
Typical inspection indications include:
- Internal cracks and discontinuities.
- Weld imperfections.
- Lack of fusion indications.
- Inclusions and voids.
- Internal delamination.
- Corrosion-related wall loss.
- Bonding defects.
- Material inconsistencies.
Detection performance depends on material characteristics, probe selection, defect orientation, calibration, component geometry, and the applicable inspection procedure.
A-Scan Signal Analysis
The D-70 provides real-time A-scan visualization for examining ultrasonic signal responses.
The display presents signal amplitude against sound travel time or calibrated distance, allowing operators to identify relevant echoes and evaluate their positions.
The instrument updates its A-scan display at approximately 60 Hz, providing responsive signal monitoring during inspection.
DAC – Distance Amplitude Correction
Distance Amplitude Correction is a standard feature of the Masterscan D-70.
DAC establishes a reference curve using known reflectors at different sound paths. This allows inspectors to compare indications at different distances while accounting for ultrasonic signal attenuation and beam characteristics.
The function is particularly useful in weld inspection and conventional ultrasonic examination where consistent amplitude evaluation is required.
TCG – Time Corrected Gain
Time Corrected Gain is available as an optional software feature.
TCG adjusts signal amplification according to sound travel time to compensate for amplitude reduction at increasing inspection distances.
This enables more consistent evaluation of ultrasonic indications across different sound paths.
AVG/DGS – Defect Sizing
The optional AVG/DGS function supports reflector evaluation using the relationship between distance, gain, and reference reflector characteristics.
It helps inspectors estimate equivalent reflector size and compare indications against established ultrasonic reference criteria.
Active Edge Technology
Active Edge is a specialized Sonatest feature designed to improve near-surface resolution.
It helps distinguish signals located close to the entry surface, where conventional ultrasonic inspection may experience limitations caused by the initial pulse and transducer characteristics.
Automatic Gain Control (AGC)
The integrated AGC function automatically adjusts signal amplification to position the signal within a predefined display range.
According to the manufacturer’s specification, AGC sets the Gate 1 signal between 10% and 90% of full screen height, with a stated tolerance range of 5% to 20%.
Measurement Modes and Inspection Capabilities
The Masterscan D-70 provides seven measurement modes for different ultrasonic examination requirements.
Mode 1 – Signal Monitoring
Provides signal monitoring without displaying measurement values. Gate alarms can remain active for inspection monitoring.
Mode 2 – Depth and Amplitude Measurement
Measures the depth and amplitude of the first signal detected within the selected gate.
Mode 3 – Echo-to-Echo Measurement
Evaluates the distance between two ultrasonic echoes, supporting thickness and sound-path measurements.
Mode 4 – Trigonometric Angle-Beam Measurement
Provides calculated inspection information, including:
- Beam path.
- Surface distance.
- Indication depth.
- Probe index offset (X-offset).
- Echo amplitude.
- Half-skip indication.
- Curved surface correction for applicable convex and concave geometries.
Mode 5 – Gate-to-Gate Measurement
Calculates the distance between signals detected in the two measurement gates.
Mode 6 – Flank-to-Flank Measurement
Determines the distance between selected signal flanks for specialized ultrasonic measurement applications.
Mode 7 – Beam Angle Measurement
Calculates the ultrasonic beam angle using beam-path information, reference-hole radius, and hole-center depth.
Ultrasonic Probe and Signal Processing Flexibility
The D-70 supports different ultrasonic transducer configurations and inspection techniques.
Supported Inspection Techniques
- Pulse-echo flaw detection.
- Transmit-receive flaw detection.
- Through-transmission testing.
- Angle-beam shear-wave testing.
- Surface-wave inspection.
- Crack diffraction measurement.
- Creep-wave inspection.
- Pitch-and-catch testing.
- Contact inspection.
- Immersion inspection.
- Single-transducer thickness measurement.
- Dual-transducer thickness measurement.
Selectable Frequency Filters
The Masterscan D-70 incorporates eight selectable frequency filter bands to accommodate different probe frequencies and inspection conditions.
These include low-frequency filters for specialized applications and higher-frequency bands for more conventional ultrasonic examination.
An additional tuned low-frequency preamplifier is available with the Dryscan option.
Advanced Display and Operator Interface
The D-70 features a color transflective VGA display designed to provide clear ultrasonic signal presentation in different working environments.
Display Features
- Adjustable screen brightness.
- Nine selectable color palettes.
- Black-on-white LCD emulation mode.
- Real-time A-scan visualization.
- Waveform freeze.
- Waveform smoothing.
- Reference waveform comparison.
- Active Peak Memory.
- Configurable measurement display.
Reference Waveform
The reference waveform function displays a previously stored A-scan in a different color from the active signal.
This allows operators to compare current inspection signals against previously recorded reference data.
Active Peak Memory
Active Peak Memory retains previous A-scan traces on the display while presenting the current waveform in a separate color.
This feature helps operators examine echo-dynamic behavior and monitor signal changes during probe movement.
Inspection Data Management with UT-Lity Software
The Sonatest D-70 works with UT-Lity software to support inspection data storage, organization, and reporting.
The standard UT-Lity software supplied with the instrument provides tools for managing inspection records on the instrument and a connected computer.
Data Management Capabilities
- Transfer stored inspection files to a PC.
- View and manage calibration settings.
- Review stored A-scans.
- Manage B-scan data.
- Organize thickness logging records.
- Create customized inspection report templates.
- Export thickness data to spreadsheets.
- Generate printable PDF reports.
- Update supported instrument firmware.
UT-Lity Pro
UT-Lity Pro is an advanced software option designed to support more comprehensive inspection planning and corrosion monitoring workflows.
Its capabilities include:
- Creating inspection plans.
- Managing inspection grids.
- Adding location notes and labels.
- Importing historical thickness readings.
- Organizing asset inspection records.
- Exporting inspection information to spreadsheets and maintenance databases.
Industrial Applications
1. Welding Inspection
The Masterscan D-70 supports conventional ultrasonic examination of welded joints in industrial fabrication, pipelines, structural steel, and pressure equipment.
Its DAC functionality and optional advanced sizing tools help inspectors evaluate ultrasonic indications according to applicable inspection procedures.
2. Oil and Gas Industry
The instrument can be used for inspecting pipelines, pressure vessels, storage tanks, and associated metallic infrastructure.
Its thickness logging and corrosion monitoring options are particularly useful for asset integrity programs.
3. Corrosion Detection
The D-70 supports ultrasonic thickness measurement and optional corrosion data logging.
These functions help maintenance personnel monitor material wall thickness and identify areas requiring further investigation.
4. Power Generation
Suitable applications include the inspection of selected turbine components, boiler equipment, metallic structures, and power generation infrastructure.
The manufacturer also identifies EMAT-related power generation applications within the Masterscan D-70 product range.
5. Composite Inspection
The instrument can support ultrasonic examination of suitable composite materials for internal discontinuities and bonding-related indications.
6. Forgings and Castings
Ultrasonic testing can help identify internal discontinuities, inclusions, and structural inconsistencies in suitable forged and cast components.
7. Aerospace and Transportation
The D-70 can support selected conventional ultrasonic inspection tasks involving structural components, manufactured parts, and bonded materials.
Actual application suitability must be established through the relevant inspection procedure and equipment configuration.
Technical Specifications
| Specification | Details |
|---|---|
| Brand | Sonatest |
| Model | Masterscan D-70 |
| Equipment Type | Digital Ultrasonic Flaw Detector |
| Testing Technology | Ultrasonic Non-Destructive Testing (NDT) |
| Measurement Modes | 7 Modes |
| Test Range | 0–20,000 mm (Steel) |
| Material Velocity | 256–16,000 m/s |
| Gain Range | 0–110 dB |
| Pulser Voltage | 100–450 V |
| Pulser Type | Negative Spike and Square Wave |
| Pulse Width | 30–2,500 ns |
| Pulse Repetition Frequency | 5 Hz–5 kHz |
| Frequency Filters | 8 Selectable Bands |
| Frequency Range | 100 kHz–33 MHz |
| Probe Modes | Pulse-Echo, Transmit-Receive, Pitch-Catch |
| Damping | 50 / 400 Ω |
| Rectification | RF, Full Wave, Positive / Negative Half-Wave |
| Measurement Gates | 2 Independent Gates |
| DAC | Standard |
| TCG | Optional |
| DGS / AVG | Optional |
| Split DAC | Optional |
| Active Edge Technology | Supported |
| Display | Color Transflective VGA TFT |
| Display Resolution | 640 × 480 Pixels |
| Screen Update Rate | 60 Hz |
| Color Palettes | 9 |
| Internal Memory | 4 GB |
| Connectivity | USB, BNC / LEMO |
| Data Management Software | UT-Lity |
| Battery Type | Rechargeable Lithium-Ion |
| Battery Operating Time | Up to 10.5 Hours |
| Operating Temperature | -10°C to +55°C |
| Protection Rating | Designed to Meet IP67 |
| Dimensions | 172 × 238 × 70 mm |
| Weight | Approximately 1.7 kg (with battery) |
| Calibration Standard | EN 12668-1:2010 |
- Battery life: The current Sonatest product page specifies up to 10.5 hours, while earlier technical literature reports approximately 12 hours.
- Frequency filter: The official technical brochure specifies a wideband filter of 1.6–22 MHz. Some third-party listings report 1.6–33 MHz; the applicable filter range should be confirmed for the supplied revision.
- IP protection: IP67 is identified in published technical brochure documentation as a design target. Confirm the applicable enclosure rating and certification before relying on it for immersion or wet-site use.
- Software options: Split DAC, TCG, AVG/DGS, AWS, API, BEA, IFT, corrosion logging, B-Chart, and Dryscan are not all standard features.
- Connector configuration: BNC and LEMO transducer connections are available depending on the factory configuration.









There are no reviews yet.